NJZ-1900B W-CDMA UE Tester

  • Measuring Instrument

Print

Discontinued on October 31, 2007

Spec.

Measurement

Power measurement

Range -60 to +36dBm
Resolution 0.1dB
Accuracy +36 to 0dBm ±1dB, 0 to -40dBm ±2dB, -40 to -60dBm ±3dB

Frequency measurement

Range ±500Hz
Resolution 1Hz
Accuracy ±83Hz

EVM measurement

Range 0 to 20%
Resolution 0.1%
Accuracy ±8.3%

BER measurement

Range 1 x 10-3 to 1 x 10-5

Current measurement

Range 0 to 1000mA
Resolution 1mA
Accuracy 0 to 100mA ±3mA
100 to 1000mA ±30mA

Transceiver

Frequency Transmit: 2110MHz - 2170MHz, Receiving: 1920MHz - 1980MHz
Number of Carrier 1
Step of Frequency 200kHz
Frequency Accuracy Based on reference oscillator
Output Level -50dBm to -110dBm
Output Level Accuracy -50 to -57.9dBm ± 3dB
-58 to -68.9dBm ± 2dB
-69 to -97.9dBm ± 3dB
-98 to -108.9dBm ± 2dB
-109 to -110dBm ± 3dB

Others

Power Supply 100 to 240VAC (50/60Hz)
Dimension 350 mm (W) x 165 mm(H) x 400 mm(D) approx.
Weight Less than 15kg

* Specifications may be subject to change without notice.